體外診斷儀器 Beckman

AU480 AU480

With a throughput of 400 photometric tests per hour (up to 800 with electrolytes) and an on board capacity of 63 different analytes the AU480 Chemistry System is the ideal main analyzer for small to medium sized laboratories.

    With a throughput of 400 photometric tests per hour (up to 800 with electrolytes) and an on board capacity of 63 different analytes the AU480 Chemistry System is the ideal main analyzer for small to medium sized laboratories. The AU480 Chemistry System can also fit as a special chemistry or STAT analyzer in large laboratories as well.

    Analytical system

    Fully automated, random access chemistry system with STAT capability

    Analytical principle

    Spectrophotometry and potentiometry

    Analytical types

    Endpoint, rate, fixed point and indirect ISE

    Analytical methods

    Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE

    Simultaneously processed analytes

    60 photometric tests + 3 ISE, 120 pre-programmed onboard tests

    Throughput

    400 photometric tests/hour; maximum of 800 with ISE

    ISE throughput

    200 samples/hr.Maximum test/hr 600 if ISE only

    Sample types

    Serum, plasma, urine, other

    Sample feeder

    Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 80 samples; continuous loading

    Sample tubes

    Primary and secondary tubes; diameter from 11.5 to 16 mm; height from 55 to 102 mm, nestled micro sample cups

    STAT samples

    Up to 22 positions for STAT samples via cooled turntable. Easy to operate interrupts between tests Auto repeat run capability

    Sample volume

    1–25 μl in 0.1 μl steps (1–25 μl for repeats)

    Reagent capacity

    76 positions for (R1+R2, detergent position). Holds 15, 30 and 60ml bottles

    Reagent storage

    Refrigerated (4 °C–12 °C

    Reagent volume

    R1: 10–250 μl; R2: 10–250 μl; (in 1 μl increment)

    Total reaction volume

    90–350 μl

    Reaction cuvette

    Permanent glass cuvettes

    Reaction time

    Up to 8 minutes, 37.5 seconds

    Reaction incubation

    37 °C, dry bath

    Photometry system

    Direct assay through the reaction cuvette (0 – 3.0 OD) mono and bichromatic measurements possible

    Wavelength

    13 different wavelengths between 340 – 800 nm

    Calibration

    Auto calibration, cooled calibrator positions; master calibration established by two-dimensional barcode

    Quality control

    Auto QC, cooled QC positions

    Rerun

    1. Auto rerun and manual rerun available
    2. Automatic sample increase, decrease or normal repeat
    3. Reflex testing

    Sample Pre Dilution

    Dilution ratio: Fixed to 3, 5, 10, 15, 20, 25, 50, 76 or 100

    Test requisition

    Individual and profile test requisition via on-line, mouse, function keys and touch screen

    Sample integrity

    Lipaemia, Haemolysis and Icterus analysis. Sample clot detection and probe crash protection

    Online

    RS232C. Full uni- and bidirectional communication possible

    Software

    Windows XP®

    Data Storage

    Sample number 100,000 samples. Reaction Data 200,000 tests

    Dimensions (W x H x D) mm

    ANL: 1,450 x 1,205 x 760

    Power supply

    100–240v; 60 Hz/ < 3.5 kVA