體外診斷儀器 Beckman

PK7300 PK7300

The Beckman Coulter PK7300 Automated Pretransfusion Blood Testing System builds on the industry established reliability and productivity of its predecessors and offers the very latest in technological advances.

    The Beckman Coulter PK7300 Automated Pretransfusion Blood Testing System builds on the industry established reliability and productivity of its predecessors and offers the very latest in technological advances.

    Electrical Requirements

     

    Electrical Consumption

    3.0 KVA maximum 

    Current

    15 ~ 30 Amp (with UPS) 

    Circuit

    Dedicated and Noise free 

    Voltage

    110/208/220/230/240 VAC (±10%)  

    Single Phase Frequency

    50/60 Hz (±1 Hz) 

    Ground Requirement

    <100 ohms 

    Location

    Power cable length 10 m (30 ft)

     

     

    Water Requirements Type

     

    Deionized Supply

    Continuous Flow 

    Resistivity

    >0.5 Mega Ohms 

    Mechanical Filtration

    <0.5 μm (glycerol free) filter at discharge of deionizer 

    Consumption

    60 liters/hour Max, instantaneous demand 3.5 liter/minute 

    Pressure

    0.49x105 - 3.99x105 pascal (7.1-56.8 PSI) 

    Location

    Shut off valve within 10 m (30 ft) 

    Tubing Diameter

    12 mm (ID) x 18 mm (OD) 

    Connection

    ½ inch barbed hose fitting

     

     

    Drain Requirements 

     

    Gravity

    Hazardous waste floor drain Maximum height - 1.5 m (5 ft) Maximum distance from analyzer - 10 m (30 ft) 

    Tubing Diameter

    15 mm (ID) x 22 mm (OD)

     

     

    Environmental Requirements

     

    Average Heat Output

    7200 KJ/H (682 BTU) Max Ambient 

    Temperature

    18 - 30° C (fluctuations during measurement shall be within +/-2°C) Ambient 

    Humidity

    20% - 80% Relative Humidity 

    Noise Output (Decibels)

    Max 65 dB or less

     

     

    General Characteristics

     

    Analytical method

    Agglutination method on terraced microplates 

    Channels

    12 Channels

    Throughput

    300 samples/hour with 4 dilution cups;  

     

    266 samples/hour with 5 dilution cups 

    Sample Capacity

    Racks with 10 samples each (barcodes on primary tubes) Capacity of 12 racks or 120 samples; continuous addition 

    Sample Tube Size

    In primary or secondary tubes; Diameter - 12 or 15 mm; height between 75 and 100 mm 

    Sample

    Plasma; Serum; Red Blood Cells; Red Blood Cell Suspension 

    Reagent Tray

    Up to 16 reagents can be loaded 

    Reaction Vessel

    Terraced microplates 

    Reaction Time

    20 - 60 minutes 

    Assays

    ABO Grouping, Rh typing, Phenotyping  (Rh, Kell and others), Irregular antibody  screening (with saline and enzyme  method), Infectious Disease screening  (TP, CMV and others)* 

    Sample Barcode

    NW-7; CODE39; CODE128; ISBT- CODE128; and 2 of 5 Interleaved